StudentShare
Contact Us
Sign In / Sign Up for FREE
Search
Go to advanced search...
Free

Acquiring Images with Three Distinct Forms of Contrast Using Sequences of Spin Echo Pulse - Assignment Example

Cite this document
Summary
"Acquiring Images with Three Distinct Forms of Contrast Using Sequences of Spin Echo Pulse" paper describes the ways of acquiring images having the three discussed dissimilar kinds of contrast, in addition to T2 contrast, in addition to having sequences of gradient reverberation pulse…
Download full paper File format: .doc, available for editing
GRAB THE BEST PAPER97% of users find it useful

Extract of sample "Acquiring Images with Three Distinct Forms of Contrast Using Sequences of Spin Echo Pulse"

MERS 7004 ASS2 Name: Course: Professor: Institution: City & State: Date: MERS 7004 ASS2 Q1. A. Acquiring images with three distinct forms of contrast using sequences of spin echo pulse A spin echo element is usually. The spin echo not only produces T1 and T2, but also protons images that are density weighted. The pulse is employed in refocusing the T2* decomposition. T1 weighted image The time of T1 relaxation for each tissue’s TI weighted image is different in characterization. Relaxation time of T1 could be described as the period needed to make sure that the remaining magnetization gets back to its original point along the longitudinal axis following the  RF pulsation. In getting the weighted image of T1, the disparities in time of relaxation of T1 require demonstration among the tissues. This can be attained by choosing a TR that is short; this makes sure that magnetization in the transverse direction does not entirely recover along the Z-axis. Water’s T1 relaxation time is longer than fat’s. As a result, after each ensuing  RF pulse, the water gains less longitudinal magnetization than the fat. Consequently, on application of the subsequent RF pulse, the transverse magnetization of the fat will be more, and that is why it produces signals, which are high, inside the receiving coils. Since water’s time of relaxation of T1 is long, it experiences slight longitudinal magnetization, as a result, it appears darker. T1 recovery is experienced due to the spins energy transfer to the environs after the  RF pulsation is applied. Increase in TR causes a high signal in the weighted image of T1 as a result of water’s and fat’s full recovery; therefore, generating a poor contrast. In general, to reduce the effects of T2 and acquire the T1-weighted image, the TE and TR ought to be shorter. The TR controls the T1 image. TR (repetition time) occurs between the original 90o RF pulse and the subsequent RF pulse. TE (echo time) occurs between the core of the obtained echo and the 90o RF pulse. T2 weighted image Exploitation of the T2 leads different tissues relaxation time developing the T2 weighted image. The relaxation time of T2 could be defined as the period needed to ensure that the remaining magnetization decomposes along the transverse plane subsequent to the 90 RF pulsations. In obtaining the weighted image of T2, the disparities in the time of relaxation of T2 have emphasis involving tissues. Through selection of a TE, which is long and allows additional time for decomposition involving water and fat, the T2 is easily acquired. Decay of the T2 depends on the relaxation or interaction of spin to spin. This leads to generation of the transverse spin phase coherence loss. As a result, the weighted image of T2 is achieved through the extended TE. T2 time of fat is short because of its hydrogen rapidly exchanging energy. Consequently, magnetization in the transverse direction is small, thus producing a signal that is low inside the receiving coils. Alternatively, water’s time of T2 is long (exchange of energy is slow). As a consequence, a large quantity of magnetization in the transverse direction generates a signal that is high inside the receiving coils. Water’s appearance on the image of T2 is bright. As noted by Thelen et al. (2008), a small TE leads to a higher signal for all tissues. This happens because of the relative similarity in the time of relaxation of T2 between tissues. Generally, longer the TE and TR, the easier it is to acquire the image of T2. The TE controls the image of T2. The Image of Proton density The protons number inside a tissue sample defines proton density. The signal magnitude depends on the protons number within the tissue sample. A case in point would be the brain, which is filled with protons, produces signals that are high; at the same time, the cortical bones, comprising small protons, produces signals that are low. In generation of images of proton density, the T1 effects and T2 effects need reduction. This is achievable through selection of a TR, which is long so that the T1 effect reduces, and a TE that is short, so that the T2 effect reduces. B. Ways of acquiring images having the three discussed dissimilar kinds of contrast, in addition to T2* contrast, in addition to having sequences of gradient reverberation pulse In gradient images, the reverberation depends on the decay of the T2* as a result of the absence of the 180o RF pulsation. A short TR and TE, crusher gradient, rewind gradients, effects of partial echo and inconsistent pulse angles, keep the images in the conventional gradient under control. These factors permit the sequence of gradient to be lesser during time of scan in comparison to the sequence of spin echo. The effect of the incomplete echo during steady state situation occurring during the sequence of gradient allows the addition of the contrast of T2 on the image of T1. Moreover, the lack of 180o RF means that the T2* cannot be refocused. Through this fashion, the gradients of reversal not only capture but also refocus the transverse de-phasing generated as a result of application of the gradients. In the gradient echo, the signal’s intensity depends on the amount of longitudinal magnetization in the steady state. The gradient echo’s signal is lesser than that of the spin sequence. The GRE’s T1-weighted image A larger angle of the RF pulse aids in controlling the T1 image with a short TR and TE in the gradient sequence. Long T1 contrast image develops as a result of the large angle RF pulse. Consequently, the short TR trims down the long T1 elements. Additionally, a gradient of slice crushers should be included (subsequent to the obtained echo) during the sequence of GRE T1 with the intention of obliterating several transverse spins generated as a result of the preceding read gradients, which could call for refocusing via the ensuing gradients from the subsequent RF pulsations, along with being appended to the reflection as a tough upright line artifact. The artifact comes about as a consequence of the large angle; T2* is equal to or more than the TR. Subsequent to the echo, there is no usage of rewind gradients inside the image of GRE T1. Water’s appearance is dark owing to the lengthened time of recovery by T1, while the appearance of fat will be bright on account of shortening T1’s time of recovery. Normally, in attaining the image of GRE T1, TE and TR are reduced and have large angles of RF pulsation. The gradient of the crusher ought also to be appended to annihilate any incoherent transverse spins attributed to the prior gradients of read. The Weighted Image of GRE T2* An amalgamation of the in-homogeneity in the gradient sequence generates the T2 effect, as well as the magnetic field’s T2*. The T2* effect is decomposition that brings about losses in the coherence of magnetization in the transverse direction. Small RF pulsation angles are essential in establishing the weighted image of T2*. Additionally, it is necessary to have long TE to multiply T2* effects and long TR to decrease T1 effects. The use of a slice crusher gradient in removing the transverse spins artifact brought about by the prior read gradients may be essential. This artifact will continue being small notwithstanding the lack of gradient of crusher. There is no need for rewind gradients subsequent to the obtained echo. In most cases, the image of T2* calls for long TEs along with TRs that have small RF pulsation angles, and could have or not have gradients of crusher. The Image of GRE density of Proton Small RF flick angles, short TEs and long TRs have an influence on the GRE proton density image. The short TE and long TR get rid of the effects of the T2 and decrease the T1 effects. The small angle pulsation causes similar longitudinal magnetization for each and every sample T1 value. This is to say that it appears bright. As a result, the contrast characterization is produced by proton density images. Although not necessary, a crusher gradient is employed in removal of the remaining transverse artifacts brought about by the preceding read gradients. As a consequence, it is not appended to the ensuing RF pulsations. By and large, the weighted image of GRE density of proton makes use of low RF pulsation angles, short TE, and long TR having or not having gradient of slice crusher. The Image of GRE T2 No image can be referred to as a chaste image of T2 inside the gradient series. However, there exists the weighted image of T1 having the contrast the T2 that is referred to as ‘additional’ T2. This is as a result of the shortness of the TR inside the gradient series. Acquisition of the image of additional T2 could be possible with a large angle variety of the RF pulsation; the contrast of T2 is added via incomplete echoes on the image of T1. It is essential that TE and TR are short. Read recovery gradients and phase rewind are utilized subsequent to the obtained echo with the intention of refocusing the remaining transverse de-phasing brought about by the other semi-portion of the gradients of read, as well as the phase gradients. These are appended to the ensuing angle of RF pulsation. This appends the signal of T2* to the image. This T2 signal will add the effect of partial spin echo ensuing from the impure angles along with the Hahn echo to the image. Therefore, a combination of these two effects, the read rewind gradients and the phase and the effect of partial spin echo, leads to the addition of the T2-weighted image on to the weighted image of T1. By and large, short TE, TR, and large angle of RF pulse generate the GRE T2 having a T1 image. More so, the rewinding of the read recovery gradient and phase gradient leads to refocusing on the effect of partial spin echo to append T2 contrast on top of the T1 image. C. Relationship involving the phantom image attained via spin echo series having a weighting of T2 (Module 7, expansion sheet on Screen 12, Image b) and the phantom reflection obtained via a gradient echo sequence having T2 weighting. The external zone is,; the internal is, and . Inside the spin echo, water’s T2 decomposition (4000 ms) in the transverse plane is long; while gel’s T2 decay (80ms) is short. Accordingly, following the 90 RF pulse, the gel decays rapidly on the outer region instigating a low signal, and so, it appears black or dark. On the contrary, the water’s decay in the transverse plane is slow bringing about signals that are high and whose appearance is either white or bright. The image of T2 found inside the spin reverberation depends on TE Inside gradients, the TE and TR are adequately short with the intention of making it impractical to image the pure T2. In the gradient’s stable state situation, the contrast of the ‘additional’ T2 is appended on top of the weighted image of T1 as a consequence of the incorporation of the gradient of phase rewind, the effect of incomplete whirl echo (signal of T2) as well as the gradient of read recovery (which is a signal of T2*). Consequently, the reduced time of T1 of the external line has T1equaling 200 ms and T2 equaling 80 ms, which looks white. The ‘additional’ T2 goes into zones that allow existence of lengthy periods of T1 and T2; particularly the water. Q2) A. Chemical shift of the artifact and how fat-water shift could be a valuable illustration of spectral width The difference in frequency between adjoining water and fat as a result of the neighboring disparities in the chemical surroundings gives rise to a chemically-induced relocation artifact. The arrangement of electrons on the spin atom provides the most significant explanation why this occurrence takes place. The enormous density of electrons trims down the results of the proton’s outside magnetic field; a phenomenon known as shielding. Fat is made up of numerous hydrogen molecules attached to an elongated group of carbons and comprises a great density of electrons (MRES 7004 module). For this reason, the water experiences a superior effectual field of magnetism than the fat. Therefore, the fat’s recessional frequency will be slower than that of water for a similar field. These differences in frequency frequently transpire at the water - fat boundary, giving rise to fat relocating from its first position in the read. The relationship between the difference in frequency in fat and water and the magnetic field is a direct proportional one. For that reason, the chemical transfers appear black in one boundary as soon as the fat relocates further than the signal of water and bright in the other boundary once the fat relocates onto the signal of water. The water and fat separation of about 3.5 ppm occurs in an autonomous field of magnetism. The equation below gives the extend pixel transfer of the bordering fat and water:  Hence,   is gyromagnetic relation, is field of magnetism, denotes bandwidth, is the amount of read direction pixels. Moreover, the relocation in pixel is reliant on the range of pixel frequency, as well as the difference in frequency. In the following instance, the difference in frequency of 220 Hz takes place in 1.5T and the range of pixel frequency (ratio between BW and the read direction’s pixel number) is 125 Hz. In that case, the shift in pixel becomes 1.76 pixels. For this reason, the shift in pixel becomes the total of each present range of pixel frequency in the difference in frequency (Biederman et al., 2007). To decrease the chemical shift artifact, it is necessary to increase the spectral bandwidth or width. Therefore, this leads to an increase in the range of pixel frequency, giving rise to a small amount of shifts of pixel (MRES 7004). Hence, the spectral bandwidth is helpful when providing explanations and giving the quantity of artifact that would have taken place on top of the image. With a low field of magnetism, a little FOV and fat repression lessens this shift in chemical artifact. b) From the equation below,  Shift in pixel - 2 pixels; -1.5T; quantity of reading pixels - 265; what would  be?  is the amount of read direction pixels × range of pixel frequency Range of pixel frequency = difference in frequency / shift in pixel  Q3) From the  in-homogeneity and given that =. The definition of the B0 in-homogeneity is the distortion or variation in the principle field of magnetism alongside the structure resulting in swift T2*decomposition, as well as incoherent phase whirls in the transverse direction. As a result, this produces spatial or intensity distortion, or both. Distortion of intensity takes place if one position is different from the magnetic field, while the other position is identical. As a consequence, the time of T2* reduces hence resulting in offset of phases. In addition, the signal becomes low inside the field of low homogeneity. These artifacts take place inside the gradient series. Shimming enhances the magnetic field’s homogeneity. Occasionally, shimming does not succeed, most likely when the ailing person is large and at the side of the cavity, where the in-homogeneity field happens. The artifact known as Moiré fringe is the end product of the winding artifact and the in-homogeneity in the field of magnetism, which needs adding up and cancellation to generate a banding look, which multiplies in the series of gradient. In the reduction of this artifact, the application of a sequence of spine is sought as opposed to a series of gradient (Horowitz, 1995). The spatial artifact originates on or after the local alteration of the field of magnetism next to the diverse weaknesses of the dissimilar boundaries of regions. These are either natural boundaries, an example being the lung’s air-tissue border, or not natural boundaries, a case in point being braces (MRES 7004 module). This brings about T2*s that are short (offset of phase) and smaller signals in the zone of local alteration fields. This artifact is clearly evident during high TE, as well as sequences of gradient. A spin echo is the substitute sequence if one’s intention is reducing spatial gradient. The artifact can be reduced by short TEs. The Effect of Partial volume The artifact caused by partial volume is established when a huge dimension of the voxel holds a blend of water and fat. As a result of the frequency difference of water and fat, the fat protons are in phase with the protons in the water at specific instances thus resulting in high signals in the reverse orientation therefore moving out of phase during other instances. This causes the signals to work against one another and their appearance in the pixels at the fat-water interface is dark. This difference in frequency involving the fat and water, in addition to modification of TE ascertains the spins’ alignment. Consequently, the quantity of the intensity the signal can be established. The artifact of partial volume takes place in both the phase and read directions. As a matter of course, it is poorer in the gradient series. The  de-phasing of the spins in the transverse plane, which is a T2, leads to the in-homogeneity and the partial volume artifacts decomposition and are larger in the series of gradient. This happens because the T2 de-phasing is not refocused by the gradient, thus, the artifacts’ signal in the image becomes higher. Nevertheless, the T2 de-phasing is refocused by the sequence of spin echo; therefore, the artifacts’ signal in the image appearing in the phase reduces. References Biederman, R.W. et al. (2007). The Cardiovascular MRI Tutorial: Lectures and Learning. New York: Lippincott Williams & Wilkins. Horowitz, A. L. (1995) MRI physics for radiologists: a visual approach. New York: Springer. Thelen, M et al. (2008) Cardiac Imaging: A Multimodality Approach. Stuttgart: Thieme Medical Publishers. MRES 7004 modules. (2011). Image sequences, gradient echo imaging, static artefacts, motion artifacts, construction and application. UQ.     Read More

As a consequence, a large quantity of magnetization in the transverse direction generates a signal that is high inside the receiving coils. Water’s appearance on the image of T2 is bright. As noted by Thelen et al. (2008), a small TE leads to a higher signal for all tissues. This happens because of the relative similarity in the time of relaxation of T2 between tissues. Generally, longer the TE and TR, the easier it is to acquire the image of T2. The TE controls the image of T2. The Image of Proton density The protons number inside a tissue sample defines proton density.

The signal magnitude depends on the protons number within the tissue sample. A case in point would be the brain, which is filled with protons, produces signals that are high; at the same time, the cortical bones, comprising small protons, produces signals that are low. In generation of images of proton density, the T1 effects and T2 effects need reduction. This is achievable through selection of a TR, which is long so that the T1 effect reduces, and a TE that is short, so that the T2 effect reduces. B. Ways of acquiring images having the three discussed dissimilar kinds of contrast, in addition to T2* contrast, in addition to having sequences of gradient reverberation pulse In gradient images, the reverberation depends on the decay of the T2* as a result of the absence of the 180o RF pulsation.

A short TR and TE, crusher gradient, rewind gradients, effects of partial echo and inconsistent pulse angles, keep the images in the conventional gradient under control. These factors permit the sequence of gradient to be lesser during time of scan in comparison to the sequence of spin echo. The effect of the incomplete echo during steady state situation occurring during the sequence of gradient allows the addition of the contrast of T2 on the image of T1. Moreover, the lack of 180o RF means that the T2* cannot be refocused.

Through this fashion, the gradients of reversal not only capture but also refocus the transverse de-phasing generated as a result of application of the gradients. In the gradient echo, the signal’s intensity depends on the amount of longitudinal magnetization in the steady state. The gradient echo’s signal is lesser than that of the spin sequence. The GRE’s T1-weighted image A larger angle of the RF pulse aids in controlling the T1 image with a short TR and TE in the gradient sequence.

Long T1 contrast image develops as a result of the large angle RF pulse. Consequently, the short TR trims down the long T1 elements. Additionally, a gradient of slice crushers should be included (subsequent to the obtained echo) during the sequence of GRE T1 with the intention of obliterating several transverse spins generated as a result of the preceding read gradients, which could call for refocusing via the ensuing gradients from the subsequent RF pulsations, along with being appended to the reflection as a tough upright line artifact.

The artifact comes about as a consequence of the large angle; T2* is equal to or more than the TR. Subsequent to the echo, there is no usage of rewind gradients inside the image of GRE T1. Water’s appearance is dark owing to the lengthened time of recovery by T1, while the appearance of fat will be bright on account of shortening T1’s time of recovery. Normally, in attaining the image of GRE T1, TE and TR are reduced and have large angles of RF pulsation. The gradient of the crusher ought also to be appended to annihilate any incoherent transverse spins attributed to the prior gradients of read.

The Weighted Image of GRE T2* An amalgamation of the in-homogeneity in the gradient sequence generates the T2 effect, as well as the magnetic field’s T2*. The T2* effect is decomposition that brings about losses in the coherence of magnetization in the transverse direction. Small RF pulsation angles are essential in establishing the weighted image of T2*. Additionally, it is necessary to have long TE to multiply T2* effects and long TR to decrease T1 effects.

Read More
Cite this document
  • APA
  • MLA
  • CHICAGO
(MERs 7004 ass2 Assignment Example | Topics and Well Written Essays - 2750 words, n.d.)
MERs 7004 ass2 Assignment Example | Topics and Well Written Essays - 2750 words. https://studentshare.org/physics/2094249-acquiring-images-with-three-distinct-forms-of-contrast-using-sequences-of-spin-echo-pulse
(MERs 7004 Ass2 Assignment Example | Topics and Well Written Essays - 2750 Words)
MERs 7004 Ass2 Assignment Example | Topics and Well Written Essays - 2750 Words. https://studentshare.org/physics/2094249-acquiring-images-with-three-distinct-forms-of-contrast-using-sequences-of-spin-echo-pulse.
“MERs 7004 Ass2 Assignment Example | Topics and Well Written Essays - 2750 Words”. https://studentshare.org/physics/2094249-acquiring-images-with-three-distinct-forms-of-contrast-using-sequences-of-spin-echo-pulse.
  • Cited: 0 times

CHECK THESE SAMPLES OF Acquiring Images with Three Distinct Forms of Contrast Using Sequences of Spin Echo Pulse

The Concept and Use of Echo Planar Imaging (EPI)

spin echo (SE) sequences are the most former MR sequences (in fact earlier to imaging) (Westbrook, 2009).... he most commonly used sequence is known as spin echo.... It is characterized by the initial application of a radio-frequency pulse of 90 degrees, followed by one more in front of 180 degrees, then double the time between these two pulses a signal or echo from stimulated tissue is successively applied with several pulse sequences of 90 and 180 degrees, each of which produces an echo which will form the radio wave which provides molecular information....
18 Pages (4500 words) Essay

Image contrast is normally described as proton density: T1-weighted or T2-weighted

a) Describe how images with the three different types of contrast are acquired with spin echo pulse sequences.... b) Describe how images proton density, T1-weighted or T2-weighted contrast, plus T2* contrast, are acquired with gradient echo pulse sequences.... The flip angle of the gradient echo pulse sequence is adjusted over a range of close to 80m degrees to get a larger tip angle in order to acquire more T1 weighting to the formed image; this will also give a smaller tip angle to provide more T2 and T2* weighting to the so formed images....
3 Pages (750 words) Essay

Magnetic Resonance Imaging Safety Protocol

In 1995, a patient's finger was severely burnt requiring a skin graft due to the negligent attachment of a pulse oximeter (Ng et al, 2003, p.... The project aimed to classify and to evaluate the value of the distinct professional perspectives which are pertinent to MRI safety Detail was allocated to the assimilation of empirical and professional and professional research on MRI safety....
40 Pages (10000 words) Essay

Sensitivity of Immediate and Delayed Gadolinium-Enhanced MRI

Sensitivity of Immediate and Delayed Gadolinium-Enhanced MRI after Injection of Gadolinium contrast of Detecting Multiple Sclerosis Lesion Name Institution INTRODUCTION The change in time has been accompanied by the emergence of various developments, including the discovery of the medical complications.... These complexities are further exacerbated by the fact that the magnetic resonance imaging is dependent on the use of the contrast agents, which further introduces other parameters that are so crucial to the diagnostic procedures that they cannot be overlooked....
23 Pages (5750 words) Dissertation

The Basic Principles Of MR Image Production

The application of a radio frequency pulse at the Larmor frequency causes a change in the distribution of spins with respect to their energy state and precessional phase coherence (Rodr´ıguez, 2003).... Magnetism results from moving electric charges or intrinsic spin moments of electrons or nuclei.... spin is a quantum mechanical property.... This means that the available spin energy levels are constrained to specific, discrete values....
10 Pages (2500 words) Essay

Magnetic Resonance Imaging Technique

The electromagnetic field has the appropriate frequency termed as resonance frequency; the protons in the magnetic field absorb and flip the spin.... The activity produces a rotating magnetic field, which is detectable using a scanner.... When one is using a powerful magnetic field of a scanner, the overall magnetic moment of different protons aligns themselves in the direction of the field.... As a result, this relaxation, radio frequency signals arise; these can be measured using receiver coils....
8 Pages (2000 words) Essay

BLIP Echo Planar Imaging Method

This paper ''BLIP Echo Planar Imaging Method'' tells that FSE (Fast spin echo), EPI is an imaging method that reduces the imaging time as it acquires multiple phase encoding steps in each repetition time (TR).... PI has three distinct characteristics in terms of speed, the variety of contrast, and flexibility in available resolution for small imaging structures like the pituitary gland.... In contrast, flexibility can be used to produce an image similar in standards to that of an SE image with the same TE and TR by using an excitation pulse in the front part followed by an 1800 pulse (T2* dephasing) to create the first echo....
10 Pages (2500 words) Report

Advantages of Higher-Field Tesla Scanners

The RF wavelength (λ) is proportional to the speed of light in the medium C medium, and inversely proportional to RF frequency f:λ αThe varying fluid content of body voxels causes variation in radio frequency which consequently allows users to acquire images with spatial resolutions and greater coverage than in the case of lower field scanners (Fschbach et al, 2004).... The push for high-field MRI is wholly tied to the benefits of potentially higher signal-to-noise ration, contrast and spectral resolution for certain specific applications....
9 Pages (2250 words) Essay
sponsored ads
We use cookies to create the best experience for you. Keep on browsing if you are OK with that, or find out how to manage cookies.
Contact Us